LIMATHERM W…N-EXI Themocouple

Mineral insulated metal sheathed insert is used as a replaceable element of temperature sensors. Insert design enables its installation in the connection head and pressing insert tip to the bottom of the sensor thermowell.

  • for intrinsically safe (Exi) temperature sensors produced by Limatherm Sensor acc. to ATEX
  • sheath material: acid-resistant steel 1.4541
			
Warning: Undefined variable $term_list in /home/users/o/olil/domains/dolcetrade.com/wp-content/themes/olil2/template-parts/post/content-listing-with-gallery.php on line 76
int(8764) NULL

Made in Poland. brand LIMATHERM, model W...N-EXI

International delivery to Europe, Asia and Africa.
Our warehouses are located in the UK, UAE and Senegal.

price by request

Available

Available

Advantages

Only the best I&C brands

Our company is a global supplier of control and measuring instruments from well-known world brands

Advantages

Discounts for organisations

We sell high-tech equipment at very low prices. We provide really advantageous conditions for partners

Advantages

Fast delivery across the Europe & Asia

We take care of all logistics and customs clearance of I&C solutions from Europe, USA, China and India

Advantages

Professional advice on device selection

Our specialists are ready to advise you on any technical question. We can find an analogue of any device

Technical specifications of W…N-EXI

Measuring range / sensing element:

  • (-40÷1200) ºC / N / class 2

Version I, pipe execution:

  • material: steel 1.4541, sheath diameter [mm]: 6, 8
  • operating temperature: max. 600°C
  • standard sheath lengths Lw[mm]:
    • 115, 175, 245, 375, 525[mm] forø6,
    • 495, 705, 995, 1395, 1995[mm] forø8, or other*
  • Lmax. [mm]= 1500 forø6
  • Lmax. [mm]= 2000 forø8

Version II, mineral insulated execution: max. operating temperature:

  • TC (N) material INCONEL
  • length L[mm]: acc. to requirements
ø3 ø4,5 ø6 ø8
450°C 550°C 700°C
900°C 1000°C 1200°C 1200°C
400°C 600°C
Load More Items

Options:

  • Temperature transmitter application: ATEX approved temperature transmitter with standard 4÷20mA output signal and with the HART communication protocol can be mounted in the connection head. Transmitter installation is carried out directly on the measuring insert (in place of a terminal block) or in the high cover connection head (solution used to enable installation of two transmitters).
  • Non-standard design: Immersion length and the measuring insert parameters can be customized per client request.

Response time to temperature change:

Thermowell diameter
[mm
Response time [s]
ø6 t0,5= 12
t0,9= 55
ø8 t0,5= 20
t0,9= 85
Load More Items

How to buy W…N-EXI

Ordering example:  APW2K/3–5000–1–SO–FlexTop2211-Ex– (0÷800)°C–Exi

User Guide and Instruction manual for LIMATHERM W...N-EXI

You can request the necessary certificates, documentation, instructions and user manuals for LIMATHERM W...N-EXI using the special form.

You can buy LIMATHERM W...N-EXI in UK or Europe at a favourable price in «DOLCHETRADE» by leaving a request on the website, by phone +441225535040 or by email . LIMATHERM W...N-EXI price are subject to change depending on exchange rates. We deliver instruments by transport companies in the United Kingdom and Europe.