Features and Benefits
▪ Detection efficiency at low particle size:
▪ D50 = 10 nm ± 1.0 nm
▪ D90 < 20 nm
▪ Extended concentration range up to 100,000 particles/cm³ without dilution (for dilution, see ‘Accessories’)
▪ Data rate up to 50 Hz – capture highly dynamic processes
▪ Integrate data directly into a network, or export from software (auto-export available)
▪ Water-removal system – compatible with high-humidity environments
▪ Improved diagnostics with Pulse Height Monitoring
▪ Fast response to rapid changes in aerosol concentration (T10-90 < 1 sec)
▪ Automatically shuts off of flow when inlet is blocked
Applications
▪ Air quality monitoring for particle number concentration (EN 16976:2024*)
▪ Atmospheric monitoring for particle size distribution (CEN/TS 17434, if part of 3938W50-CEN10)
▪ Health effect studies
▪ Basic aerosol research
▪ Combustion
Particle Size Range | 10 nm minimum detectable particle size (D50), verified with monodisperse sucrose particles Efficiency of 90% at Dp < 20 nm > 3 um max. detectable particle size | |
Particle Concentration Range | Up to 100,000 (1×105) particles/cm³ Single particle counting mode with continuous live-time coincidence correction | |
Particle Concentration Accuracy | ±5% at <100,000 particles/cm³ | |
False Background Counts | <0.001 particles/cm³ based on 12-hour average | |
Response Time | (Response time is described as a percentage of a concentration step change) < 1 second for 90% to 10% (T10-90, T90-10) ~2 seconds for 0 to 95% (T95) |
|
Flow System | 1.0 ± 0.05 L/min inlet and counting flow (volumetric)
Requires external vacuum source capable of 60 kPa (18 in Hg) minimum gauge (below atmospheric pressure); Pump model 3032-EC, listed under ‘Accessories’, meets this requirement |
|
Liquid System | Butanol (n-Butyl alcohol, not included) used as working fluid
Internal water removal pump to remove condensate; beneficial in humid environments. |
|
Data Storage | Internal memory lasts for ~ 1 year of data at 50 Hz data rate | |
Communication Interfaces | Ethernet port for remote connection: 8-wire RJ-45 jack, 10/100 BASE-T, TCP/IP).
Configurable for automated (DHCP) or manual network settings. USB type C to connect CPC directly to computer (cable included) USB type B for external memory drives; a Wifi adapter can be used Pulse output: BNC connector, TTL level pulse, nominally 350 ns wide Embedded touch-display |
|
Ambient Operating Conditions | Temperature
Humidity Pressure |
10 to 35°C (50 to 95°F)
0 to 90% RH, non-condensing 75 to 105 kPa (0.75 to 1.05 atm) |
Electrical | 100 to 240 VAC, 50/60 Hz, 200 W maximum | |
Accessories | 3750200
3333-10
RHT3000
375X-2LBOTTLE
AIM11CPCMONTRIAL
IMCPCMONITORING |
Sampling System for Atmospheric Particles
Aerosol Diluter
Relative Humidity and Temperature Sensor
2 L fill bottle
CPC Monitoring Software Trial: permits Aerosol Instrument Manager (monitoring license) |
Dimensions (H x W x D) | 27.5 cm x 18.3 cm x 29.9 cm (10.83 in. x 7.21 in. x 11.76 in.), not including fill bottle and bracket |
|
Weight | ~6.6 kg (~14.6 lbs) |
You can request the necessary certificates, documentation, instructions and user manuals for TSI 3750-10 using the special form.
You can buy TSI 3750-10 in UK or Europe at a favourable price in «DOLCHETRADE» by leaving a request on the website, by phone +441225535040 or by email dolce@dolcetrade.com. TSI 3750-10 price are subject to change depending on exchange rates. We deliver instruments by transport companies in the United Kingdom and Europe.