+441225535040

Mon-Fri: 8:00 - 18:00 UK

Email:

NAGMAN R40 Sub-Zero Temperature Dry Block Calibrator

  • Portable
  • Cost Effective
  • Fast heating & Cooling
  • PID Temperature Control
  • Switch Test Facility
			
Warning: Undefined variable $term_list in /home/users/o/olil/domains/dolcetrade.com/wp-content/themes/olil2/template-parts/post/content-listing-with-gallery.php on line 76
int(6961) NULL

Made in India. brand NAGMAN, model R40

International delivery to Europe, Asia and Africa.
Our warehouses are located in the UK, UAE and Senegal.

price by request

Available

Available

Advantages

Only the best I&C brands

Our company is a global supplier of control and measuring instruments from well-known world brands

Advantages

Discounts for organisations

We sell high-tech equipment at very low prices. We provide really advantageous conditions for partners

Advantages

Fast delivery across the Europe & Asia

We take care of all logistics and customs clearance of I&C solutions from Europe, USA, China and India

Advantages

Professional advice on device selection

Our specialists are ready to advise you on any technical question. We can find an analogue of any device

Description of NAGMAN R40

Ideal for Field & Lab Calibration purpose.

  • Basic Instrument
  • Cable for Thermostat Switch Test
  • Mains Cable
  • Insertion Tube (to suit 1/4“ & ½” probe)
  • Tool for Insertion Tube
  • Spare Fuses
  • Traceable Calibration certificate
  • Instruction Manual
  • Carrying Case

Technical specifications of NAGMAN R40

  • Range: -10°C to 123°C
  • Resolution: 0.1 °C
  • Accuracy: ±0.5 °C
  • Stability: ±0.1 °C
  • Stabilization Time: 15 Minutes
  • Temperature Readout: °C / °F Selectable
  • Well Diameter: 19 mm
  • Immersion Depth: 100 mm
  • Heating Time: 10 Minutes (from Ambient to max.) (approx.)
  • Cooling Time: 20 Minutes (from Ambient to min.) (approx.)
  • Interchangeable Thermowells: Single Hole to suit 1/4″ & ½” Probes – Standard
  • Switch Test Facility / Supply: Provided / 5V DC (open)
  • Power Supply: 230 V AC / 50 Hz
  • Computer Interface
  • “Caltemp” Calibration Software
  • Additional Thermowells / Insertion Tubes (to choose from) :
    Single hole standard sizes to suit : 1/8”, 1/4”, 3/8”, 1/2” & 3, 4, 5, 6, 7, 8, 10, 12, 13, 15 & 17 mm probes
    Multihole (Typical) : (1×6 + 1×8), (3×6), (1×10 + 1×4) mm
  • Power Supply 110 V AC, 60 Hz
  • Calibration Certificates are issued in Accordance with our Scope as granted by NABL per ISO/IEC 17025:2017 Standards